FDA-Device2020-02-05Class II

Bard Conquest 40 PTA Dilation Catheter, 8mm X 20mm x 75cm (CQF7582), 8mm x 40mm x 75cm (CQF7584); Recommended Guidewire .035"; Recommended Introducer 6F; Nominal Pressure 8atm, Rated Burst Pressure 40atm; Syringe inflation 1cc; UDI: (01)00801741060533(17)220228(10)REDQ3444

Bard Peripheral Vascular Inc
Hazard

The catheter packaging incorrectly listed the balloon size as 8mm X 20mm instead of the intended size 8mm X 40mm. If undetected, the user will advance the longer than expected balloon to the lesion. The potential exists that upon inflation, non-diseased vessel, kidney or sections of the AV dialysis fistulae may be dilated, potentially causing a varying degree of injury. This may require additional medical or surgical intervention.

Sold states
US: NJ, NY, OH, PA, VA, SC, GA, AR, TX, MA, NC, TN, LA, AZ, MS, FL, AL,MD, CA, HI, MO, CO, WI, DC, KY, IL, MN, IN, KS, OK, CT, MT, WA, ND, IA, WA, WV, PR OUS: None
Affected count
286 Units
Manufactured in
1625 W 3rd St Bldg 1, N/A, Tempe, AZ, United States
Products
Bard Conquest 40 PTA Dilation Catheter, 8mm X 20mm x 75cm (CQF7582), 8mm x 40mm x 75cm (CQF7584); Recommended Guidewire .035"; Recommended Introducer 6F; Nominal Pressure 8atm, Rated Burst Pressure 40atm; Syringe inflation 1cc; UDI: (01)00801741060533(17)220228(10)REDQ3444

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0917-2020