FDA-Device2021-02-03Class II
Brain Heart Infusion Agar with 6 ug/ml Vancomycin Cat no: G14 Lot no: 472698 Manufactured: 2020-10-29 Expires: 2020-12-28 Container Type: 15x100mm plate, 18 ml
Hardy Diagnostics
Hazard
Complaints were received reporting of Staphylococcus aureus (ATCC 29213) breakthrough when the strain should have been inhibited.
Sold states
U.S.: CA, UT, MI, Puerto Rico, IN, CT, AZ, GA, MA, TN, WA O.U.S. None
Affected count
820 units (82 PK10)
Manufactured in
1430 W McCoy Ln, Santa Maria, CA, United States
Products
Brain Heart Infusion Agar with 6 ug/ml Vancomycin Cat no: G14 Lot no: 472698 Manufactured: 2020-10-29 Expires: 2020-12-28 Container Type: 15x100mm plate, 18 ml
Official notice