FDA-Device2014-02-12Class II

B. Braun AccuFlo Elastomeric Infusion Pump 200ML/H 100 ML Generally used for delivery of antibiotic and local anesthesia medications.

B. Braun Medical, Inc.
Hazard

Internal testing of two lots of the AccuFlo CT-2000-100 Elastomeric Infusion Pump identified flow rates which are outside the specified infusion time and may result in overly rapid infusion of medications.

Sold states
US Distribution one importer/wholesale distributor in MO.
Affected count
30, 240 units
Manufactured in
901 Marcon Blvd, Allentown, PA, United States
Products
B. Braun AccuFlo Elastomeric Infusion Pump 200ML/H 100 ML Generally used for delivery of antibiotic and local anesthesia medications.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0921-2014