FDA-Device2020-02-12Class II

Giraffe Blue Spot PT LiteTM Phototherapy System, Model Numbers M1224917, M1231533, M1225025, M1224961, M1224958, M1224957, M1224959, M1224963, M1225773, M1225066, M1224960, M1225059, M1225023, M1225060, M1225063, M1225064, M1225024, M1231622, M1236303, M1238489, M1238490, M1238492, M1238493, M1238495, M1238499, M1238503, M1238505, M1238507, M1238508, 2063860-001, 2063862-001, 2063870-001, 2063873-001, 2063874-001, 2071393-001, 2080916-001, 2089424-001

Lumitex Inc
Hazard

There is a potential for some systems built between 2014 and Jan 2018 to have a light output falling below the recommended minimum output. This problem has been associated with exposure of the light pipe to elevated temperatures, which is contra-indicated in the Operation, Maintenance, and Service Manual.

Sold states
The products were distributed to the following US states: WI.
Affected count
6837
Manufactured in
8300 Dow Cir, N/A, Strongsville, OH, United States
Products
Giraffe Blue Spot PT LiteTM Phototherapy System, Model Numbers M1224917, M1231533, M1225025, M1224961, M1224958, M1224957, M1224959, M1224963, M1225773, M1225066, M1224960, M1225059, M1225023, M1225060, M1225063, M1225064, M1225024, M1231622, M1236303, M1238489, M1238490, M1238492, M1238493, M1238495, M1238499, M1238503, M1238505, M1238507, M1238508, 2063860-001, 2063862-001, 2063870-001, 2063873-001, 2063874-001, 2071393-001, 2080916-001, 2089424-001

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0923-2020