FDA-Device2015-01-14Class II

CelluTome Epidermal Harvesting System Harvester (5.0 CM) STERILE EO Rx Only Manufactured for: KCI USA, Inc. Part Number CT-H50 or CT-H25 The CelluTome Epidermal Harvesting System is intended to reproducibly cut a thin skin graft for autologous skin grafting.

KCI USA, INC.
Hazard

KCI has received reports that, in a small number of cases, the CelluTome Harvester (a component of the CelluTome Epidermal Harvesting System), blades have the potential to drift during shipping and could prematurely protrude into openings of harvester top plate.

Sold states
Worldwide Distribution - USA (Nationwide) and Internationally to Australia, Austria, Canada, France, Germany, Ireland, Italy, Isreal, Netherlands, South Africa, Spain, Sweden, Switerland, Turkey, UAE, and United Kingdom.
Affected count
1,139/5 Packs
Manufactured in
12930 IH 10 West, N/A, San Antonio, TX, United States
Products
CelluTome Epidermal Harvesting System Harvester (5.0 CM) STERILE EO Rx Only Manufactured for: KCI USA, Inc. Part Number CT-H50 or CT-H25 The CelluTome Epidermal Harvesting System is intended to reproducibly cut a thin skin graft for autologous skin grafting.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0933-2015