FDA-Device2022-04-20Class II

Trumpf Mount Clamp Assembly (table adapter component), part #118704-000, a component of the IMRIS HFD100 head fixation device, models #119629-000 and #119630-000, which are used with ORT400 operating room tables.

Deerfield Imaging, Inc.
Hazard

Reports have been received that after normal or routine lateral forces are applied to the head fixation device, the split block clamp may deflect (rotationally around the axis) and not return to within acceptable distance from its prior position after the force is removed. In order to reduce deflection at the table mount location, users have tightened the hand knob screw to a degree that it is then difficult to untighten it later in the clinical workflow. Rotational movement at the table attachment point could result in HFD movement going unnoticed, which could result in a situation where surgical or medical intervention is necessary, a situation where navigation needs to be reestablished, or a situation resulting in extended time under anesthesia for the patient. It may also be difficult to loosen the mount clamp component after use, potentially leading to ergonomic injury or a workaround during equipment teardown.

Sold states
Worldwide distribution - US Nationwide distribution in the states of AL, FL, GA, MA, MN, NC, NY, OH, and TX. The countries of Sweden and Switzerland.
Affected count
24
Manufactured in
5101 Shady Oak Rd S, N/A, Minnetonka, MN, United States
Products
Trumpf Mount Clamp Assembly (table adapter component), part #118704-000, a component of the IMRIS HFD100 head fixation device, models #119629-000 and #119630-000, which are used with ORT400 operating room tables.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0934-2022

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief