FDA-Device2021-02-10Class II
ALung Technologies Hemolung Respiratory Assist System (RAS) with CME BodyGuard 323 Infusion Pump(EUA) - Product Usage: an off-the-shelf infusion pump and administration set to provide a 30 mL/hr continuous saline flush to the integrated blood pump bearing.
Alung Technologies Inc
Hazard
CME America BodyGuard 323 Infusion Pumps (recalled by manufacturer) is a component to the Hemolung for EUA, and the infusion pump may have a delivery inaccuracy
Sold states
US Nationwide distribution.
Affected count
101 units
Manufactured in
2500 Jane St Ste 1, Pittsburgh, PA, United States
Products
ALung Technologies Hemolung Respiratory Assist System (RAS) with CME BodyGuard 323 Infusion Pump(EUA) - Product Usage: an off-the-shelf infusion pump and administration set to provide a 30 mL/hr continuous saline flush to the integrated blood pump bearing.