FDA-Device2021-02-10Class II
Tourni-Cot - Universal, UDI 00855364004085, Model Number TCU, Non-pneumatic Tourniquet - Product Usage: The device is intended to be used in surgery on fingers or toes to provide a blood-less field.
Mar-Med Co
Hazard
A failure can occur if the user attempts to remove the device by partially snipping the edge and pulling on it. The device may tear into two pieces with a portion remaining in place, which can continue to constrict the vessels. This type of failure can result in digit ischemia.
Sold states
Worldwide distribution - US Nationwide distribution and the countries of Australia, Canada, France, Germany, Greece, Italy, Japan, Netherlands, Singapore, Spain, Sweden, Switzerland, and United Kingdom.
Affected count
52,000
Manufactured in
333 Fuller Ave Ne, Grand Rapids, MI, United States
Products
Tourni-Cot - Universal, UDI 00855364004085, Model Number TCU, Non-pneumatic Tourniquet - Product Usage: The device is intended to be used in surgery on fingers or toes to provide a blood-less field.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0945-2021Don't want to check this manually?