FDA-Device2013-03-20Class II
TotalCare SpO2RT2 bed. Product Usage: Bed, flotation therapy, powered bed may be used in a variety of settings including, but not limited to, acute care, including critical care, step down/progressive care, medical/surgical, high acuity sub-acute care, post anesthesia care unit (PACU), and sections of the emergency department (ED). The TotalCare SpO2RT¿ 2 Bed System is capable of being used with a broad patient population as determined appropriate by the caregiver or institution.
Hill-Rom, Inc.
Hazard
In certain situations a software problem with the TotalCare SpO2RT¿ 2 bed may occur that can lead to inadvertent loss of sufficient surface cushion pressure to adequately support the patient. This may increase the likelihood of skin irritation for patients at higher risk of developing pressure-related skin conditions. A correction to the software program is currently being developed and tested,
Sold states
USA Nationwide Distribution including the state of Canada
Affected count
309
Manufactured in
1069 State Route 46 East, N/A, Batesville, IN, United States
Products
TotalCare SpO2RT2 bed. Product Usage: Bed, flotation therapy, powered bed may be used in a variety of settings including, but not limited to, acute care, including critical care, step down/progressive care, medical/surgical, high acuity sub-acute care, post anesthesia care unit (PACU), and sections of the emergency department (ED). The TotalCare SpO2RT¿ 2 Bed System is capable of being used with a broad patient population as determined appropriate by the caregiver or institution.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0949-2013Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief