FDA-Device2023-01-18Class II

IMMULITE 2000 Thyroglobulin (OUS). For in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers for the quantitative measurement of thyroglobulin in serum or heparinized plasma, as an aid in monitoring patients who have undergone thyroidectomy. Catalog # Catalog # L2KTY2 Siemens Material Number (SMN): 10381648

Siemens Healthcare Diagnostics, Inc.
Hazard

Potential for Functional Sensitivity to not meet Instructions For Use (IFU) claims with the kit lots; control level one may result outside of published ranges; when control results are in range, users may observe increased imprecision with low level patient samples

Sold states
Nationwide Foreign: Argentina¿¿¿¿¿¿ Australia¿¿¿¿¿¿ Austria¿¿¿¿¿¿¿¿ Bahrain¿¿¿¿¿¿¿¿ Bangladesh¿¿¿¿¿ Belgium¿¿¿¿¿¿¿¿ Bolivia¿¿¿¿¿¿¿¿ Bosnia¿Herzeg.¿ Brazil¿¿¿¿¿¿¿¿¿ Bulgaria¿¿¿¿¿¿¿ Canada¿¿¿¿¿¿¿¿¿ Chile¿¿¿¿¿¿¿¿¿¿ Colombia¿¿¿¿¿¿¿ Costa¿Rica¿¿¿¿¿ Croatia¿¿¿¿¿¿¿¿ Cyprus¿¿¿¿¿¿¿¿¿ Czech¿Republic¿ Dominican¿Rep.¿ Ecuador¿¿¿¿¿¿¿¿ Egypt¿¿¿¿¿¿¿¿¿¿ Estonia¿¿¿¿¿¿¿¿ Finland¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Georgia¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Honduras¿¿¿¿¿¿¿ Hong¿Kong¿¿¿¿¿¿ Hungary¿¿¿¿¿¿¿¿ India¿¿¿¿¿¿¿¿¿¿ Indonesia¿¿¿¿¿¿ Iran¿¿¿¿¿¿¿¿¿¿¿ Iraq¿¿¿¿¿¿¿¿¿¿¿ Ireland¿¿¿¿¿¿¿¿ Israel¿¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Jordan¿¿¿¿¿¿¿¿¿ KA Kazakhstan¿¿¿¿¿ Kosovo¿¿¿¿¿¿¿¿¿ Kyrgyzstan¿¿¿¿¿ Latvia¿¿¿¿¿¿¿¿¿ Lithuania¿¿¿¿¿¿ Macedonia¿¿¿¿¿¿ Malaysia¿¿¿¿¿¿¿ Mexico¿¿¿¿¿¿¿¿¿ Morocco¿¿¿¿¿¿¿¿ Nepal¿¿¿¿¿¿¿¿¿¿ Netherlands¿¿¿¿ P.R.¿China¿¿¿¿¿ Pakistan¿¿¿¿¿¿¿ Panama¿¿¿¿¿¿¿¿¿ Paraguay¿¿¿¿¿¿¿ Peru¿¿¿¿¿¿¿¿¿¿¿ Philippines¿¿¿¿ Poland¿¿¿¿¿¿¿¿¿ Portugal¿¿¿¿¿¿¿ Republic¿Korea¿ Romania¿¿¿¿¿¿¿¿ Russian¿Fed.¿¿¿ Saudi¿Arabia¿¿¿ Serbia¿¿¿¿¿¿¿¿¿ Singapore¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ Slovenia¿¿¿¿¿¿¿ South¿Africa¿¿¿ Spain¿¿¿¿¿¿¿¿¿¿ Sweden¿¿¿¿¿¿¿¿¿ Switzerland¿¿¿¿ Taiwan¿¿¿¿¿¿¿¿¿ Trinidad,Tobago Tunisia¿¿¿¿¿¿¿¿ Turkey¿¿¿¿¿¿¿¿¿ Turkmenistan¿¿¿ U.A.E.¿¿¿¿¿¿¿¿¿ Ukraine¿¿¿¿¿¿¿¿ United¿Kingdom¿ Uruguay¿¿¿¿¿¿¿¿ Uzbekistan¿¿¿¿¿ Vietnam¿¿¿¿¿¿¿¿
Affected count
7,543 kits
Manufactured in
511 Benedict Ave, Tarrytown, NY, United States
Products
IMMULITE 2000 Thyroglobulin (OUS). For in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers for the quantitative measurement of thyroglobulin in serum or heparinized plasma, as an aid in monitoring patients who have undergone thyroidectomy. Catalog # Catalog # L2KTY2 Siemens Material Number (SMN): 10381648

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0950-2023