FDA-Device2013-03-27Class II

Howmedica Osteonics Corp. Stryker Orthopaedics Specialty Triathlon Left and Right A//P Sizer Adjustment Housing PER File K2401. For use in knee prosthesis implantation. NON-STERILE

Stryker Howmedica Osteonics Corp.
Hazard

The standard Triathlon Femoral Stylus does not fit into the Specialty Sizers and the "R" and "L" markings on the Right Sizer are reversed.

Sold states
1 Branch (NJ) with one consignee (DE).
Affected count
4 units (total of 2 sets)
Manufactured in
325 Corporate Drive, N/A, Mahwah, NJ, United States
Products
Howmedica Osteonics Corp. Stryker Orthopaedics Specialty Triathlon Left and Right A//P Sizer Adjustment Housing PER File K2401. For use in knee prosthesis implantation. NON-STERILE

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0964-2013