FDA-Device2020-02-12Class II
C-Flex Double Pigtail Ureteral Stent Set, Global Product No. G14364 G14365 G14367
Cook Inc.
Hazard
The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
Sold states
Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.
Affected count
13388 total
Manufactured in
750 N Daniels Way, N/A, Bloomington, IN, United States
Products
C-Flex Double Pigtail Ureteral Stent Set, Global Product No. G14364 G14365 G14367
Official notice