FDA-Device2024-02-14Class II
Ingenia Ambition X (Product Number 781356) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system
Philips North America
Hazard
Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.
Sold states
Worldwide - US Nationwide distribution in the states of FL and NY. The countries of Denmark, Norway, and Sweden.
Affected count
10
Manufactured in
222 Jacobs St, Cambridge, MA, United States
Products
Ingenia Ambition X (Product Number 781356) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0975-2024Don't want to check this manually?