FDA-Device2013-04-03Class II
Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 20 Count/40 Capacity; Sterile; Removable Numbered Foam Block; Adhesive Area (Cat. No. 4FA40SSA) Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 20 Count/40 Capacity; Sterile; Removable Numbered Foam Block; Magnetic Area (Cat. No. 4FM40SSA) Product Usage: Disposal system for counting of used surgical blades and needles.
Cardinal Health
Hazard
Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.
Sold states
Worldwide Distribution - USA (Nationwide) and the countried of Australia, Canada and Japan.
Affected count
Cat. No. 4FA40SSA: 96; Cat. No. 4FM40SSA: 32,592
Manufactured in
1430 Waukegan Rd, Mc Gaw Park, IL, United States
Products
Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 20 Count/40 Capacity; Sterile; Removable Numbered Foam Block; Adhesive Area (Cat. No. 4FA40SSA) Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 20 Count/40 Capacity; Sterile; Removable Numbered Foam Block; Magnetic Area (Cat. No. 4FM40SSA) Product Usage: Disposal system for counting of used surgical blades and needles.
Official notice