FDA-Device2013-04-03Class II
Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 15 Count/30 Capacity; Sterile; Removable Numbered Foam Block; Magnetic Area (Cat. No. 4FS30SSA) Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 15 Count/30 Capacity; Sterile; Removable Numbered Foam Block; Double Foam Strips (Cat. No. 4FS31SSA) Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 15 Count/30 Capacity; Sterile; Double Numbered Magnets (Cat. No. 4MM30SSA) Product Usage: Disposal system for counting
Cardinal Health
Hazard
Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.
Sold states
Worldwide Distribution - USA (Nationwide) and the countried of Australia, Canada and Japan.
Affected count
Cat. No. 4FS30SSA: 7488; Cat. No. 4FS31SSA: 48; Cat. No. 4MM30SSA: 6336
Manufactured in
1430 Waukegan Rd, Mc Gaw Park, IL, United States
Products
Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 15 Count/30 Capacity; Sterile; Removable Numbered Foam Block; Magnetic Area (Cat. No. 4FS30SSA) Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 15 Count/30 Capacity; Sterile; Removable Numbered Foam Block; Double Foam Strips (Cat. No. 4FS31SSA) Needle Counters with Blade Disarmer; 11.5 x 6.5 x 3.3 cm; 15 Count/30 Capacity; Sterile; Double Numbered Magnets (Cat. No. 4MM30SSA) Product Usage: Disposal system for counting of used surgical blades and needles.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0980-2013Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief