FDA-Device2020-02-12Class II
Genius 3 Tympanic Thermometer- electric thermometer Item Code: 303013
Cardinal Health 200, LLC
Hazard
The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (¿ 0.2¿C for Genius 2 and ¿ 0.3¿C for Genius 3 thermometers). The measurement readings drift upwards over time, which means that the thermometers could exceed the upper stated accuracy tolerance of +0.2¿C for Genius 2 or +0.3¿C for Genius 3.
Sold states
Worldwide Distribution - US Nationwide and Foreign in the countries of: Australia Austria Belgium Canada Canary Islands Chile Costa Rica Croatia Cyprus Czech Republic Denmark Egypt Estonia Finland France French Guiana French Polynesia Gabon Germany Gibraltar Greece Guadeloupe Hong Kong Iran, Islamic Republic Of Ireland Israel Italy Kuwait Liechtenstein Luxembourg Martinique Mayotte Mexico Netherlands New Caledonia New Zealand Norway Panama Philippines Poland Portugal Puerto Rico Qatar Reunion Romania San Marino Saudi Arabia Singapore Slovakia Slovenia South Africa South Korea Spain Swaziland Sweden Switzerland Taiwan Thailand Turkey United Arab Emirates United Kingdom
Affected count
7425 US; OUS 214278
Manufactured in
15 Hampshire St Bldg 5, N/A, Mansfield, MA, United States
Products
Genius 3 Tympanic Thermometer- electric thermometer Item Code: 303013
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0991-2020Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief