FDA-Device2022-05-04Class II
NeuMoDx Cartridge-For in vitro diagnostic use with NeuMoOx Molecular Systems with either of the following assays: NeuMoDx SARS CoV-2 Test Strip (REF 300800) NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip (REF 300900) NeuMoDx Laboratory Developed Tests (LDTs) for SARS-CoV-2 RNA detection Note: Results for targets other than SARS-CoV-2 are not impacted by this Urgent Medical Device Correction Ref. Number: 100100
NeuMoDx Molecular Inc
Hazard
Potential for false positive results for SARS-CoV-2 in identified lots of cartridges, when used in conjunction with assays containing SARS-CoV-2 targets (NeuMoDx" SARS-CoV-2 Test Strip Cat #300800 and NeuMoDx" Flu A-B/RSV/SARS-CoV-2 Vantage Assay Cat #300900) result incorrect management of patients
Sold states
Worldwide distribution - US Nationwide distribution in the states of OH, MO, VA, MI, MI, FL, MD, IA, NY, MO and the countries of Italy, Germany, Switzerland, Sweden, Netherlands, Finland, Spain, United Kingdom, Belgium, Austria, Hong Kong, France, Denmark, Slovenia.
Affected count
773 (72 US and 701 ROW) cases (48 cartridges/case)
Manufactured in
1250 Eisenhower Pl, Ann Arbor, MI, United States
Products
NeuMoDx Cartridge-For in vitro diagnostic use with NeuMoOx Molecular Systems with either of the following assays: NeuMoDx SARS CoV-2 Test Strip (REF 300800) NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip (REF 300900) NeuMoDx Laboratory Developed Tests (LDTs) for SARS-CoV-2 RNA detection Note: Results for targets other than SARS-CoV-2 are not impacted by this Urgent Medical Device Correction Ref. Number: 100100
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1000-2022Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief