FDA-Device2017-01-18Class II
FV-RF (Clinix RF Plus) Product Usage: A Stationary X-ray System, a device intended for radiographic/fluoroscopic examinations of various anatomical regions. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.
Philips Medical Systems (Cleveland) Inc
Hazard
There's a potential for liquid penetration into the back side of the front control panel of the X-ray diagnostic table, which can cause a short circuit of the terminals of the tilt operation and cause an unintentional tilt of the table.
Sold states
US Nationwide distribution in the following states: AL, CA, FL, GA, HI, IL, IN, KS, LA, MA, MI, MS, NJ, NY, OH, PA, SC, VA, WI, WV, & WY.
Affected count
41
Manufactured in
595 Miner Rd, Cleveland, OH, United States
Products
FV-RF (Clinix RF Plus) Product Usage: A Stationary X-ray System, a device intended for radiographic/fluoroscopic examinations of various anatomical regions. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1005-2017Don't want to check this manually?