FDA-Device2014-02-26Class II

VERSA-FX KEYLESS 90D X 8 and X 11, VERSA-FX KEYLESS 95D X 8, X 11, and X 14, VERSA-FX KEYLESS 140D X 4 and X 5, VERSA-FX KEYLESS 135D X 4 and X 5, VERSA-FX SHORT TUBE, 130D, 135D and 140D, VERSA-FX STANDARD TUBE, 1 and VERSA-FX SPRCNDLR TUBE, 9 Item codes starting with 001980, 001981, 0011990 and 0011991

Zimmer, Inc.
Hazard

Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.

Sold states
Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand, and United Kingdom.
Affected count
7,044,680 total
Manufactured in
345 E Main St, Warsaw, IN, United States
Products
VERSA-FX KEYLESS 90D X 8 and X 11, VERSA-FX KEYLESS 95D X 8, X 11, and X 14, VERSA-FX KEYLESS 140D X 4 and X 5, VERSA-FX KEYLESS 135D X 4 and X 5, VERSA-FX SHORT TUBE, 130D, 135D and 140D, VERSA-FX STANDARD TUBE, 1 and VERSA-FX SPRCNDLR TUBE, 9 Item codes starting with 001980, 001981, 0011990 and 0011991

Official notice