FDA-Device2023-02-01Class II

Single-Use Packs, REF: OPO80

Johnson & Johnson Surgical Vision, Inc.
Hazard

Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

Sold states
US: FL, MS, NY, TN, TX, IL, MN, OH. OUS: Belgium, China, Estonia, France, Germany, Indonesia, Italy, Japan, Malaysia, Netherlands, Portugal, Sri Lanka, Switzerland, Thailand, United Kingdom
Affected count
384
Manufactured in
31 Technology Dr Ste 200, N/A, Irvine, CA, United States
Products
Single-Use Packs, REF: OPO80