FDA-Device2024-02-14Class II
Water Filter MAJ-2318. This MAJ-2318 water filter has been designed to be used with Olympus endoscope reprocessor for prevent contamination of the rinse water. The MAJ-2318 water filter is compatible with OER-ELITE and OER-MINI.
Hazard
Water filter may have been assembled incorrectly, resulting in unfiltered tap water being used for reprocessing. This may expose a patient to a potentially contaminated device, presenting a risk of infection or sepsis.
Sold states
Domestic US distribution nationwide.
Affected count
877 MAJ-2318, 119 OER-Elite, 3 OER-MINI
Manufactured in
3500 Corporate Pkwy, Center Valley, PA, United States
Products
Water Filter MAJ-2318. This MAJ-2318 water filter has been designed to be used with Olympus endoscope reprocessor for prevent contamination of the rinse water. The MAJ-2318 water filter is compatible with OER-ELITE and OER-MINI.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1009-2024More Olympus Corporation of the Americas recalls
- FDA-Device2026-04-08Olympus SOLTIVE Premium laser system recalled for defective power supply
- FDA-Device2026-04-08Olympus SOLTIVE Pro SuperPulsed Laser System recalled for power supply defect
- FDA-Device2026-04-01Olympus PKS Cutting Forceps recalled for defective welds causing jaw breakage
- FDA-Device2026-04-01Olympus MAJ-1443 and MAJ-1444 endoscope suction valves no longer compatible with OER-Pro and OER-Elite reprocessors
- FDA-Device2026-04-01Olympus PKS Cutting Forceps recalled for defective welds causing jaw breakage
Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief