TRILOGY ACET SHELL 36MM O TRILOGY ACET SHELL 38MM O TRILOGY ACET SHELL 40MM O TRILOGY ACET SHELL 42MM O TRILOGY ACET SHELL 44MM O TRILOGY ACET SHELL 46MM O TRILOGY ACET SHELL 48MM O TRILOGY ACET SHELL 50MM O TRILOGY ACET SHELL 52MM O TRILOGY ACET SHELL 54MM O TRILOGY ACET SHELL 56MM O TRILOGY ACET SHELL 58MM O TRILOGY ACET SHELL 60MM O TRILOGY ACET SHELL 62MM O TRILOGY ACET SHELL 64MM O TRILOGY ACET SHELL 66MM O TRILOGY ACET SHELL 68MM O TRILOGY ACET SHELL 70MM O TRILOGY ACET S
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1024-2014Don't want to check this manually?