FDA-Device2013-04-10Class II

Integra Ojemann Cortical Stimulator; Product / Catalogue No: OCS2. Intended for intraoperative cortical stimulation mapping procedures.

Integra Burlington MA, Inc.
Hazard

If the headphone jack is in use during a surgical procedure with the OCS2 and a non-intended voltage (such as static electricity) comes in contact with the outer case of the unit, that voltage could be transmitted to the patient and could be a potential source of injury.

Sold states
Worldwide Distribution-USA (nationwide) and the countries of Austria, Belgium, Czech Republic, Belgium, Denmark, Spain, Great Britain, Hungary, Ireland, Italy, Netherlands, Poland, Morocco, Netherlands, Poland, Portugal, Saudia Arabia, Sweden, Turkey, and South Africa.
Affected count
371 units
Manufactured in
22 Terry Ave, N/A, Burlington, MA, United States