FDA-Device2015-02-11Class II

Animas Vibe Insulin Infusion Pump. This product is indicated for continuous subcutaneous infusion of insulin for the treatment of diabetes and has a continuous glucose monitoring feature.

Animas Corporation
Hazard

Calibration factors in the pump overwritten during a programming step. The force sensor could send a lower signal value to the pump processor, with loss of prime warnings, occlusion alarms and the pump unable to detect a cartridge during the prime sequence. Field action initiated 8/29/2011.

Sold states
No US distribution, Distributors are located in France, Germany, Sweden and United Kingdom.
Affected count
1235
Manufactured in
200 Lawrence Dr, N/A, West Chester, PA, United States
Products
Animas Vibe Insulin Infusion Pump. This product is indicated for continuous subcutaneous infusion of insulin for the treatment of diabetes and has a continuous glucose monitoring feature.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1034-2015