PERI SCR. 3.5MM X 10MM PERI SCR. 3.5MM X 12MM PERI SCR. 3.5MM X 14MM PERI SCR 3.5MM X 44MM STE PERI SCR. 3.5MM X 65MM PERI. SCR 4.0MM X24MM PERI. SCR 4.0MM X40MM PERI SCR. 4.0MMX14MM FULL PERI SCR. 4.0MMX32MM FULL PERI SCR. 4.0MMX38MM FULL 3.5MM CORT SCR X 10MM 3.5MM CORT SCR X 12MM 3.5MM CORT SCR X 14MM 3.5MM CORT SCR X 16MM 3.5MM CORT SCR X 18MM 3.5MM CORT SCR X 20MM 3.5MM CORT SCR X 22MM 3.5MM CORT SCR X 24MM 3.5MM CORT SCR
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1040-2014Don't want to check this manually?