FDA-Device2015-02-11Class II

Cobas b 221 (versions 5 and 6) Blood Gas and Electrolytes Analyzer

Roche Diagnostics Operations, Inc.
Hazard

Chemical component of the F3 Fluid pack and environmental bacterial contamination causes a positive bias up to 39% at 4.1 mmol/l for the glucose measurement and 30% at 4.12 mmol/l for lactate. Urea readings are unaffected.

Sold states
US Nationwide distribution.
Affected count
103
Manufactured in
9115 Hague Rd, N/A, Indianapolis, IN, United States
Products
Cobas b 221 (versions 5 and 6) Blood Gas and Electrolytes Analyzer

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1041-2015