FDA-Device2015-03-04Class II

Philips Multi Diagnost Eleva with FlatDetector

Philips Medical Systems, Inc.
Hazard

The possibility of the Reference Air Kerma Rate for low dose mode will not be reduced to 50% of the value for the normal mode, but will be approximately 70-80%, leading to a higher patient dose than expected.

Sold states
Worldwide Distribution - USA ( nationwide ) to the states of : AR, AZ, CA, CO, DE, FL, GA, HI, IL, IA, KY, KS, LA, MI, MO, MS, NC, NY, OH, PA, TN, TX, VA, VT, WA, WY and WV., and the countries of Austria, Australia, Colombia, Germany, Switzerland, Turkey, Ireland, Dubia and United Kingdom.
Affected count
238
Manufactured in
3000 Minuteman Rd, N/A, Andover, MA, United States
Products
Philips Multi Diagnost Eleva with FlatDetector

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1045-2015

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief