GE HealthCare SMV BodyTrack nuclear medicine system recalled for detector fall risk
GE HealthCare identified that certain older SMV BodyTrack nuclear medicine systems may have been moved or transported without proper detector support during their service life. This can weaken the detector mounting mechanisms. If the detector becomes loose or falls, it could cause serious injury. No injuries or detector falls have been reported so far.
From the FDA notice · Read the official notice ↗ · All GE Healthcare recalls
What to do
- If your facility operates an older SMV BodyTrack system, contact GE HealthCare immediately to verify if your unit is affected
- Do not move or relocate the system without contacting GE HealthCare for guidance
- Ensure the detector is properly secured and supported before any use
- Follow GE HealthCare's instructions for inspection and any necessary repairs
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- SMV BodyTrack, System, Tomography, Computed, Emission
- Sold in
- Nationwide
- Made for
- All ages (0 months and older)
This recall covers 4 products — see all of them.
Details
- Units affected
- 12 units
- Illnesses reported
- None reported
- Manufactured in
- 4, Hayozma St, Tirat Carmel, Israel
- Classification
- Class II
- Recall number
- Z-1046-2026
In the FDA's words
- Description
- SMV BodyTrack, System, Tomography, Computed, Emission GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems. UDI/DI Not applicable: All serial numbers in distribution Distribution: US, Brazil, Egypt, Canada, Mexico, Turkey, Italy, Sweden, Switzerland, France, Spain Chile, Germany, Iran, Greece Affected quantity: 12 units
- Hazard
- GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
4 products in this recall
GE HealthCare nuclear imaging systems recalled for detector fall risk
GE HealthCare is recalling four older nuclear medicine imaging systems, 60 units in total, that are past their end of guaranteed service. The systems may have been moved without proper detector support, which could stress the detector mounts and lead to a detector falling. No falls or injuries have been reported.
The agency lists each product separately. Check every one you might have.
- SMV DSX New Line, System, Tomography, Computed, EmissionZ-1044-2026 · 7 units
- SMV DSI, System, Tomography, Computed, EmissionZ-1045-2026 · 3 units
- SMV BodyTrack, System, Tomography, Computed, Emission (this page)Z-1046-2026 · 12 units
- MAXXUS, System, Tomography, Computed, EmissionZ-1047-2026 · 38 units