GE HealthCare SMV BodyTrack nuclear medicine system recalled for detector fall risk
Check GE HealthCare nuclear medicine system detector support
GE HealthCare identified that certain older SMV BodyTrack nuclear medicine systems may have been moved or transported without proper detector support during their service life. This can weaken the detector mounting mechanisms. If the detector becomes loose or falls, it could cause serious injury. No injuries or detector falls have been reported so far.
- If your facility operates an older SMV BodyTrack system, contact GE HealthCare immediately to verify if your unit is affected
- Do not move or relocate the system without contacting GE HealthCare for guidance
- Ensure the detector is properly secured and supported before any use
- Follow GE HealthCare's instructions for inspection and any necessary repairs
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1046-2026Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief