FDA-Device2013-04-10Class III

Siemens Dimension(R) EXL(TM) LM STM (Siemens Material Number 106393377). Intended to measure a variety of analytes in human body fluids.

Siemens Healthcare Diagnostics, Inc.
Hazard

Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined that many of these errors are due to the current limit set for System Check LOCI results. The errors have no impact on results, but they may cause customer inconvenience. Siemens h

Sold states
Worldwide Distribution.
Affected count
2207 total
Manufactured in
500 Gbc Dr Ms 514, PO BOX 6101, Newark, DE, United States
Products
Siemens Dimension(R) EXL(TM) LM STM (Siemens Material Number 106393377). Intended to measure a variety of analytes in human body fluids.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1047-2013

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief