FDA-Device2019-04-03Class II
LW Scientific ZIP IQ PCV Centrifuge, ZIC-24HD-75T3 Rev C Product Usage: used as a general purpose centrifuge to spin down blood in hematocrit tubes
LW Scientific, Inc.
Hazard
The clear, plastic, polycarbonate hematocrit rotor may break apart during use and be ejected through the plastic housing, posing a safety hazard to the technician and potentially damaging the surrounding area.
Sold states
Worldwide Distribution - US Nationwide and the countries of Philippines, Honduras, El Salvador, Australia, Guatemala, Canada, Singapore, Australia, Peru, Nicaragua, Trinidad and Tobago, and New Zealand
Affected count
142 units
Manufactured in
865 Marathon Pkwy, N/A, Lawrenceville, GA, United States
Products
LW Scientific ZIP IQ PCV Centrifuge, ZIC-24HD-75T3 Rev C Product Usage: used as a general purpose centrifuge to spin down blood in hematocrit tubes
Official notice