FDA-Device2015-02-11Class II
LINAC Accelerators: ARTISTE, ONCOR Avant Garde, ONCOR Impression Plus, ONCOR Expression, ONCOR Impression, PRIMUS HI, Mevatron M2/Primus Mid-Energy, MEVATRON PRIMART, MEVATRON KD-2, MEVATRON KDS-2, MEVATRON KDS-2, MEVATRON KD-2, MEVATRON K2, MEVATRON 6323-2, MEVATRON M2 6740, MEVATRON M2 6700, MEVATRON M2 6300, MEVATRON MDX-2, MEVATRON MD-2; The intended use of the MEVATRON, PRIMUS, ONCOR and ARTISTE MV System linear accelerator system is to deliver x-ray radiation for therapeutic
Siemens Medical Solutions USA, Inc
Hazard
A hardware update to replace the old PCB Assembly, Head Driver-G41 with PCB with overcurrent protection as a result of reports of the electric board inside the gantry showing signs of overheating, including thermal damage and smoke emission.
Sold states
Worldwide Distribution - US Nationwide including Canada and Mexico.
Affected count
310
Manufactured in
51 Valley Stream Pkwy, N/A, Malvern, PA, United States
Products
LINAC Accelerators: ARTISTE, ONCOR Avant Garde, ONCOR Impression Plus, ONCOR Expression, ONCOR Impression, PRIMUS HI, Mevatron M2/Primus Mid-Energy, MEVATRON PRIMART, MEVATRON KD-2, MEVATRON KDS-2, MEVATRON KDS-2, MEVATRON KD-2, MEVATRON K2, MEVATRON 6323-2, MEVATRON M2 6740, MEVATRON M2 6700, MEVATRON M2 6300, MEVATRON MDX-2, MEVATRON MD-2; The intended use of the MEVATRON, PRIMUS, ONCOR and ARTISTE MV System linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer. Product Usage: The intended use of the MEVATRON, PRIMUS, ONCOR and ARTISTE MV System linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1059-2015Don't want to check this manually?