FDA-Device2015-02-18Class II

CARDIOSAVE Intra-Aortic Balloon Pump and CS300 Intra-Aortic Balloon Pump; and, non-sterile, not for human use, SENSATION and SENSATION PLUS IABC demonstration units. Part Number 0998-00-0800-xx; 0998-00-3023-xx; 0998-UC-0800-xx; 0998-UC-3023-xx An electromechanical system used to inflate and deflate intra-aortic balloons.

Maquet Datascope Corp - Cardiac Assist Division
Hazard

Ferrule Cap was not removed on some fiber-optic Sensation and Sensation Plus IABC demonstration units. If fiber-optic connector is inserted into the Maquet CS300 or CARDIOSAVE IABP, cap will prevent connection of unit and may become dislodged and remain in IABPs fiber-optic receptacle.

Sold states
Worldwide Distribution
Affected count
838 units-US; 364 units - OUS
Manufactured in
1300 MacArthur Blvd., N/A, Mahwah, NJ, United States
Products
CARDIOSAVE Intra-Aortic Balloon Pump and CS300 Intra-Aortic Balloon Pump; and, non-sterile, not for human use, SENSATION and SENSATION PLUS IABC demonstration units. Part Number 0998-00-0800-xx; 0998-00-3023-xx; 0998-UC-0800-xx; 0998-UC-3023-xx An electromechanical system used to inflate and deflate intra-aortic balloons.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1062-2015