FDA-Device2013-04-17Class II
Toshiba TSX-30IC AQUlLlON ONE System, Computed Tomography X-Ray. This system supports Whole body scanning.
Toshiba American Medical Systems Inc
Hazard
Due to a problem with the control software of the x-ray high-voltage generator, if scanning is performed with a tube current less than 30mA, the control software may detect an error and perform error processing even though the x-ray output is normal.
Sold states
Distributed in the state of MD.
Affected count
1 unit
Manufactured in
2441 Michelle Dr, P.O. Box 2068, Tustin, CA, United States
Products
Toshiba TSX-30IC AQUlLlON ONE System, Computed Tomography X-Ray. This system supports Whole body scanning.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1068-2013Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief