FDA-Device2013-04-17Class II

Toshiba TSX-30IC AQUlLlON ONE System, Computed Tomography X-Ray. This system supports Whole body scanning.

Toshiba American Medical Systems Inc
Hazard

Due to a problem with the control software of the x-ray high-voltage generator, if scanning is performed with a tube current less than 30mA, the control software may detect an error and perform error processing even though the x-ray output is normal.

Sold states
Distributed in the state of MD.
Affected count
1 unit
Manufactured in
2441 Michelle Dr, P.O. Box 2068, Tustin, CA, United States
Products
Toshiba TSX-30IC AQUlLlON ONE System, Computed Tomography X-Ray. This system supports Whole body scanning.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1068-2013