FDA-Device2025-02-05Class II

muRata Vios Monitoring System Model 2050; bedside cardiac monitor, including cardiotachometer and alarm rate

Murata Vios, Inc.
Hazard

During the set-up and workflow to begin Vios monitoring, it has been noticed that patient vital signs and ECG waveforms can become distorted and will flicker or flash very fast. This issue only occurs on the Central Station Monitor (CSM) in the patient tiles view and is not seen at any other point during patient monitoring & not on the Bedside Monitor.

Sold states
Worldwide - US Nationwide distribution in the states of NY, TX, OH, MD and the country of India.
Affected count
62 units
Manufactured in
700 Commerce Dr Ste 190, Woodbury, MN, United States
Products
muRata Vios Monitoring System Model 2050; bedside cardiac monitor, including cardiotachometer and alarm rate

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1069-2025