FDA-Device2022-05-18Class II
1 Series CX phototherapy units equipped with Daavlin's ClearLink Control System software versions v3.03h and 3.04v. Models: [1 Ser CX 305-4 120-240V 50/60Hz, item # 06OS0004CX6], [1 Ser CX 311-4 120-240V 50/60Hz CE, item # 807OS0004CX5], [1 Ser CX 311-4 120-240V 50/60Hz, item # 807OS0004CX6], [1 Ser CX 350-4 120-240V 50/60Hz, item # 807OS0400CX6], [1 Ser CX UVA1-4 120-240V 50/60Hz, item # 807OS0400CX6A1]
Daavlin Distributing Company
Hazard
Software issue, resuming an interrupted treatment will result in swap of utilized calibration treatment distance 0" inches to "9. Resulting in over or under dosing.
Sold states
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, VA, WA, WI, WV and the countries of Australia, Canada.
Affected count
216
Manufactured in
205 W Bement St, N/A, Bryan, OH, United States
Products
1 Series CX phototherapy units equipped with Daavlin's ClearLink Control System software versions v3.03h and 3.04v. Models: [1 Ser CX 305-4 120-240V 50/60Hz, item # 06OS0004CX6], [1 Ser CX 311-4 120-240V 50/60Hz CE, item # 807OS0004CX5], [1 Ser CX 311-4 120-240V 50/60Hz, item # 807OS0004CX6], [1 Ser CX 350-4 120-240V 50/60Hz, item # 807OS0400CX6], [1 Ser CX UVA1-4 120-240V 50/60Hz, item # 807OS0400CX6A1]