FDA-Device2019-04-10Class II
The UniCel DxI 600 and 800 Access Immunoassay Systems, that are connected to an automation line. UniCel DxI 600 Catalog Numbers: A30260, A71460, A71461, A92060. Unicel DxI 800 Catalog Numbers: 973100, A71456, A71457, A84545, A25288, A25285.
Beckman Coulter Inc.
Hazard
When racks are front-loaded onto the sample presentation unit while samples from the automation line are being processed, a scheduling error within the software can cause mishandling of samples during the creation of an aliquot and lead to incorrect results being generated. Affected samples typically yield non-reproducible falsely decreased results by up to 100%, however the potential for falsely elevated results cannot be ruled out.
Sold states
US: NJ, OH, NY, GA, FL, TN, WI, MI, MS, CA, NC, TX, PA, DC, IN, KY, IL, WA, MN, LA, ND, ID, WV, AK, AL, MA. OUS: Argentina, Australia, Bahrain, Canada, China, Croatia, Czech Republic, Egypt, Germany, Greece, India, Italy, Malaysia, Morocco, Netherlands, Nigeria, Poland, Portugal, Qatar, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Thailand, Tunisia, United Kingdom, United States, Vietnam
Affected count
2842
Manufactured in
250 S Kraemer Blvd, N/A, Brea, CA, United States
Products
The UniCel DxI 600 and 800 Access Immunoassay Systems, that are connected to an automation line. UniCel DxI 600 Catalog Numbers: A30260, A71460, A71461, A92060. Unicel DxI 800 Catalog Numbers: 973100, A71456, A71457, A84545, A25288, A25285.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1077-2019Don't want to check this manually?