FDA-Device2013-04-17Class II
GE Healthcare, Discovery MR750w 3.0T System. Part Number: 5352293. Product Usage: It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra,dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate,blood vessels, and musculoskeletal regions of the body.
GE Healthcare, LLC
Hazard
GE Healthcare has recently become aware of a potential issue due to image flip in frequency encoding direction associated with a specific workflow of your Discovery MR750w system. Image flip issue in frequency encoding direction (including S/I, A/P and R/L) can result in incorrect localization of anatomy and pathological findings.
Sold states
Worldwide Distribution - USA (nationwide) including the states of: AZ, CA, CO, FL, IL, MD, NJ, NY, OH, OK, TX. and the countries of Australia, Belarus, Canada, France, Germany, India, Japan, Lebanon, New Zealand, Saudi Arabia, Singapore, Spain, Sweden, Switzerland and United Kingdom.
Affected count
46
Manufactured in
3000 N Grandview Blvd, Waukesha, WI, United States
Products
GE Healthcare, Discovery MR750w 3.0T System. Part Number: 5352293. Product Usage: It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra,dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate,blood vessels, and musculoskeletal regions of the body.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1100-2013Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief