Moderate riskFDA DeviceClass II

RayStation radiation therapy planning software recalled for potential dose calculation errors

RayStation radiation therapy treatment planning software (versions 14.0.0, 15.0.0, and 15.1.3) may not properly update calculated radiation doses when certain regions of interest are modified. This can occur when regions have no contours but have material overrides or are classified as bolus, fixation, or support types. Changes to these regions may not trigger the dose recalculation as intended.

From the FDA notice · Read the official notice ↗ · All Raysearch Laboratories recalls

What to do

  1. Contact RaySearch Laboratories if your facility uses affected RayStation software versions
  2. Review recent treatment plans created with versions 2023B, 2024A, or 2024A SP3
  3. Do not rely on automatic dose recalculation for regions of interest with no contours until your software is updated

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 14.0.0, 15.0.0, 15.1.3 Software Version: RayStation 2023B, RayStation 2024A, RayStation 2024A SP3 Product Description: Radiation Therapy Treatment Planning System Component: No
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 9 products — see all of them.

Details

Units affected
119
Illnesses reported
None reported
Manufactured in
Eugeniavagen 18c, Stockholm, Sweden
Classification
Class II
Recall number
Z-1106-2026

In the FDA's words

Description
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 14.0.0, 15.0.0, 15.1.3 Software Version: RayStation 2023B, RayStation 2024A, RayStation 2024A SP3 Product Description: Radiation Therapy Treatment Planning System Component: No Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended. Lot Code: UDI: 0735000201055620230630, 0735000201072320231213, 0735000201088420250129 GTIN: 07350002010556, 07350002010723, 07350002010884 Serial Numbers: 14.0.0.3338, 15.0.0.430, 15.1.3.10 Software Revisions: RayStation 2023B, RayStation 2024A, RayStation 2024A SP3 Expiration Date: 2030-02-21 Distribution: Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS Affected quantity: 119
Hazard
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

9 products in this recall

RayStation treatment planning software recalled for dose invalidation error

RaySearch Laboratories is recalling nine RayStation and RayPlan radiation therapy treatment planning software entries. Certain regions of interest with no contours, which have a material override or are of type Bolus, Fixation, or Support, may not invalidate calculated radiation dose as intended when their geometry is changed or their material override is removed. No injuries have been reported.

The agency lists each product separately. Check every one you might have.

More from Raysearch Laboratories

See all