FDA-Device2013-04-24Class II

CoaguChek XS Pro meters (catalog number 05530199160)bar code scanner. Used with the CoagCHEK XS Pro Meters to quantitatively determines prothrombin time ("PT").

Roche Diagnostics Operations, Inc.
Hazard

Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards. Errors have been confirmed with the EAN 13 barcodes which should not be used for patient or operator identification.

Sold states
Nationwide Distribution.
Affected count
4467 both devices
Manufactured in
9115 Hague Rd, N/A, Indianapolis, IN, United States
Products
CoaguChek XS Pro meters (catalog number 05530199160)bar code scanner. Used with the CoagCHEK XS Pro Meters to quantitatively determines prothrombin time ("PT").

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1114-2013