FDA-Device2017-02-08Class II
VITROS Na+ Slides, (UPN 10758750004812) For in vitro diagnostic use only. VITROS Na+ Slides quantitatively measure sodium (Na+) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
Ortho-Clinical Diagnostics
Hazard
There is a potential for biased results to be generated over the 10 day on-analyzer storage limit when using VITROS Na+ Slide cartridges warmed at room temperature and loaded onto the analyzer after using the minimum warm up requirements stated in the IFU. If Quality Control (QC) results are within acceptable limits, reported results are acceptable and were not affected by this issue.
Sold states
Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Bermuda, Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain and Poland.
Affected count
659,957 Total Units (249,570 units domestically & 410,387 units internationally)
Manufactured in
513 Technology Blvd, N/A, Rochester, NY, United States
Products
VITROS Na+ Slides, (UPN 10758750004812) For in vitro diagnostic use only. VITROS Na+ Slides quantitatively measure sodium (Na+) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.