Moderate riskFDA DeviceClass II

ICU Medical Plum Solo Precision IV Pump recalled for piggyback flush programming error

ICU Medical's Plum Solo and Duo Infusion pumps have a software issue that may prevent clinicians from programming an automatic line flush after piggyback therapy. In certain situations, the pump displays an error message that blocks the flush from being set up, even though the flush is a normal part of the treatment process.

From the FDA notice · Read the official notice ↗ · All ICU Medical recalls

What to do

  1. If you use or manage these pumps, contact ICU Medical for guidance on the affected software versions (1.0.2 for US model 400010401 or 1.0.0 for Philippines model 400011001)
  2. Do not attempt workarounds; follow your facility's protocol and ICU Medical's instructions

Is yours affected?

Compare your label or packaging with the recalled product.

Product
ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philippines)
Sold in
Nationwide
Made for
All ages (under 100 years)

This recall covers 3 products — see all of them.

Details

Illnesses reported
None reported
Manufactured in
600 N Field Dr, Lake Forest, IL, United States
Classification
Class II
Recall number
Z-1120-2026

In the FDA's words

Description
ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philippines) Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed. 400010401 (US), Software Versions: 1.0.2; 400011001 (Philippines), Software Versions: 1.0.0 Distribution: Worldwide distribution - US Nationwide and the country of Philippines.
Hazard
Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.

3 products in this recall

ICU Medical Plum infusion pumps recalled for piggyback flush error

ICU Medical is recalling three Plum infusion pump models: the Plum Duo Infusion Pump, Plum Duo Precision IV Pump, and Plum Solo Precision IV Pump. In certain cases, the pump's automatic flush feature for piggyback therapy can show a limit violation message that prevents the flush from being programmed. No injuries have been reported.

The agency lists each product separately. Check every one you might have.