FDA-Device2026-01-28Class II
ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philippines)
ICU Medical, Inc.
Hazard
Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.
Sold states
Worldwide distribution - US Nationwide and the country of Philippines.
Manufactured in
600 N Field Dr, Lake Forest, IL, United States
Products
ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philippines)
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1120-2026Don't want to check this manually?