FDA-Device2020-02-19Class II

Prometra II Programmable 20mL Pump. Catalog No. 93827 The device is an implantable infusion pump. GTIN-14 00810335020YYY (YYY= in-country language specific)

Flowonix Medical Inc
Hazard

A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system.

Sold states
Nationwide domestic distribution. Foreign distribution to United Kingdom, Republic of Ireland, Poland, Greece, Italy, Germany, Netherlands, and Belgium.
Affected count
1083 (OUS)
Manufactured in
120 Forbes Blvd Ste 170, Mansfield, MA, United States
Products
Prometra II Programmable 20mL Pump. Catalog No. 93827 The device is an implantable infusion pump. GTIN-14 00810335020YYY (YYY= in-country language specific)

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1125-2020