FDA-Device2015-02-25Class II

Restoris Partial Knee Application (PKA) RIO (TGS 2.0). For use with the Robotic Arm Interactive Orthopedic System (RIO), is indicated for use in surgical knee procedures.

Mako Surgical Corporation
Hazard

When using the MAKOplasty partial knee Arthroplasty application, the burr continues spinning outside of the stereotactic boundary and after the control switches (foot pedal and trigger) cease to be activated

Sold states
Nationwide Distribution-including the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, LA, MA, MD, MI, MO, MS, NC, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV and Hawaii.
Affected count
228 units
Manufactured in
2555 Davie Rd Ste 110, Plantation, FL, United States
Products
Restoris Partial Knee Application (PKA) RIO (TGS 2.0). For use with the Robotic Arm Interactive Orthopedic System (RIO), is indicated for use in surgical knee procedures.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1128-2015