FDA-Device2021-03-03Class II
Slush + Warmer Disc Drape; Model ORS-320 Slush + Warmer Disc Drapes - Product Usage: single-use equipment cover intended for use during various surgeries where warm irrigation, slush and/or cold solution is required.
Ecolab Inc
Hazard
Slits were identified in a low quantity of ORS-320 Slush + Warmer Disc Drapes from a single lot. A source for potential manufacturing error was identified and investigation is underway to prevent further recurrence. The slits may lead to a breach in the sterile barrier between fluid in the basin and the slush/warming equipment. This could pose a risk of infection to patients receiving warmed irrigation fluid or surgical slush if used during a procedure.
Sold states
Worldwide distribution - US Nationwide Distribution in the states of AL, AZ, CA, CO, FL, HI, ID, IL, IN, KS, LA, MA, MD, MI, MO, MT, NJ, NY, OH, OK, OR, PA, TN, TX, UT, WA. In the country of Japan.
Affected count
132 cases (3168 drapes)
Manufactured in
1 Ecolab Pl, N/A, Saint Paul, MN, United States
Products
Slush + Warmer Disc Drape; Model ORS-320 Slush + Warmer Disc Drapes - Product Usage: single-use equipment cover intended for use during various surgeries where warm irrigation, slush and/or cold solution is required.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1141-2021Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief