FDA-Device2018-03-28Class II
Wash Kit (REF 10680) Lots 16279, 17045 and 17215 Medica corporation is marketing a wash solution Chemistry Wash1 , REF # 10680 for use with Medica s EasyRA clinical chemistry analyzer. The solution is used prior to a lipase (LIP) assay if an interferant assay (TRIG)is part of the work list. The major interferant for Lipase is the Triglycerides (TRIG) assay since one of the reagents contains lipase as an active ingredient.
Medica Corporation
Hazard
The firm has initiated a field recall of the Wash1 Kit REF 10680 due to reported inconsistent Lipase values when analyzing patient samples. The firm's internal investigation identified the underlying cause to be an ineffective wash solution resulting in an interference carryover from the triglyceride (TRIG) assay if they are on the same worklist.
Sold states
Worldwide and US Distribution.
Affected count
392
Manufactured in
5 Oak Park Dr, N/A, Bedford, MA, United States
Products
Wash Kit (REF 10680) Lots 16279, 17045 and 17215 Medica corporation is marketing a wash solution Chemistry Wash1 , REF # 10680 for use with Medica s EasyRA clinical chemistry analyzer. The solution is used prior to a lipase (LIP) assay if an interferant assay (TRIG)is part of the work list. The major interferant for Lipase is the Triglycerides (TRIG) assay since one of the reagents contains lipase as an active ingredient.