FDA-Device2014-03-19Class II

Panaroma Central Station Product Usage: This device can view real time, store, print, graph, and trend patient clinical and demographic data. The device can also set independent alarm limits for data send by the bedside monitor.

Mindray DS USA, Inc. dba Mindray North America
Hazard

Mindray has identified two issues with the Panorama Central Station. Full Disclosure data may lag behind the Panorama system time clock when the Panorama Central Station is in use with Mindray¿s V-Series Monitor. Additionally, event indicators lag behind the Panorama system time clock when the Panorama Central Station is in use with a Panorama Telepack.

Sold states
Worldwide Distribution - USA Nationwide Distribution
Affected count
732 units in the US
Manufactured in
800 MacArthur Blvd, N/A, Mahwah, NJ, United States
Products
Panaroma Central Station Product Usage: This device can view real time, store, print, graph, and trend patient clinical and demographic data. The device can also set independent alarm limits for data send by the bedside monitor.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1156-2014