Moderate riskFDA DeviceClass II

Philips Patient Information Center iX software update silently changes notification settings

Philips released a software update (version 4.5.0) for the Patient Information Center iX that changes how mobile event notifications are filtered. The system makes these changes without alerting users, so staff may not realize their notification settings have been altered.

From the FDA notice · Read the official notice ↗ · All Philips recalls

What to do

  1. Review your mobile event notification filter settings after any system updates
  2. Contact Philips support if you notice unexpected changes to notification behavior
  3. Check with your clinical engineering or IT team about any recent software patches

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Patient Information Center iX; Software Version Number: 4.5.0
Sold in
Nationwide
Made for
All ages (0 months and older)

Details

Units affected
146 units (41 US, 105 OUS)
Illnesses reported
None reported
Manufactured in
222 Jacobs St, Cambridge, MA, United States
Classification
Class II
Recall number
Z-1156-2026

In the FDA's words

Description
Patient Information Center iX; Software Version Number: 4.5.0 A recent software (SW) patch modifies Mobile Event Notification filter settings without providing any indication to the user when upgrading the system. Model Numbers: (1) 866389, (2) 866390; UDI-DI: (1) 00884838127074, (2) N/A; Serial Numbers: 4Z24-5CKL-P, 0P6C-4TL3-D, 6J58-3LL6-0, 216G-7PGV-E, 0F2U-1BLC-H, 291E-7HJU-A, 1M6W-1UP5-1, 4620-0XNW-5, 4W5R-3FHA-L, 6B73-3ZKP-K, 3E79-49M4-3, 4F6Y-70HU-P, 0128-2WG5-0, 0830-3RLW-W, 097K-7LPC-5, 0D2V-1BLC-8, 120T-17M9-A, 173C-20JM-J, 190B-64MU-U, 1K2J-15LB-P, 1L3T-7KGB-E, 2K06-4ZPR-X, 3916-5DJ6-Z, 3V7H-06MT-E, 412A-2WG5-Y, 4271-2NL7-M, 4Z3L-5PP3-0, 5D0E-3EH3-3, 5E0B-3EH3-D, 5E25-14LG-9, 657X-31N5-L, 6G5U-3FHA-R, 781T-7TJA-B, 7N3U-51MJ-B, 7P31-7NPM-P, 283X-7ZPB-A, 7H77-09GU-Z, 4C59-26K6-P, 4N0L-0FKE-2, 5Y0A-02P6-L, 494H-1CJJ-3, 326G-7PGV-A, 292G-3YNZ-9, 0V2A-6DHP-0, 6354-58M8-C, 4W5B-6JK9-A, 5M7K-0VKT-T, 3401-1FHE-M, 6R30-5BNT-N, 6N2E-7YK1-1, 6R0H-6AMX-T, 107C-62PF-Y, 2X1T-7TJA-H, 1M6G-7PGV-K, 1C4Z-6KKD-H, 0B3Z-2RHR-L, 7H39-3JGK-U, 0M0V-6FKB-5, 793D-41PZ-X, 233G-46MR-V, 782Z-3RLW-D, 5N2B-5YJP-G, 0L5F-0ZNF-J, 4G4N-2LKY-K, 2M48-4XM7-Y, 133W-3UN7-B, 295N-0YG8-Y, 5D7W-00G0-1, 5R78-6CPV-0, 3C79-49M4-9, 3T24-5CKL-A, 6H74-27LY-T, 186W-1UP5-8, 0021-1DNN-P, 0247-7ELE-R, 0D5Y-32PN-A, 0J7W-31N5-2, 0R2W-1VNP-P, 0Y5Z-7FHX-N, 121L-1KG1-M, 1366-7XLV-X, 144W-7RPJ-5, 1472-5WKF-X, 1C19-1MHK-F, 1G0F-69LP-G, 1J5U-5HMV-N, 1P4F-34MB-W, 1P5C-6JK9-D, 212R-6HHJ-P, 221V-7TJA-4, 227N-7JG6-J, 232W-7DL9-G, 2A5X-2PJT-0, 2A75-27LY-3, 2F5J-0JLM-7, 2J19-1MHK-W, 2L16-59N8-M, 2M73-5WKF-U, 2N3V-6YM2-X, 2P6J-7PGV-E, 2Y5M-5AGH-E, 326W-21KC-B, 332M-47P7-Y, 3A29-0CKW-3, 3C0U-7AHH-6, 3C2U-05HF-W, 3E05-4ZPR-U, 3G6H-39G4-Y, 3H1T-22HW-5, 3H5R-5HMV-M, 3H79-09GU-5, 3J2K-2TP9-1, 3W1Y-3CH6-J, 4L7Y-00G0-F, 4M77-1HM0-G, 4N3D-20JM-K, 4R0G-60JV-1, 4V2Z-7NPM-M, 4W08-0UH7-U, 4X6C-1PJ5-J, 4Y08-0UH7-L, 500W-6WNX-A, 594D-0EG2-9, 5U0A-3WPK-C, 5W1G-50P2-N, 5Z6J-3HL1-B, 617W-31N5-F, 637V-7BG9-W, 673T-67ME-B, 686N-3NNV-6, 6D3P-4WMG-A, 6K70-3XNY-W, 6P42-35J4-V, 6U1T-75JN-Y, 771A-2JPY-0, 7F5L-73HZ-3, 7G3A-13JG-L, 7K0P-4AH2-6, 7Z2R-05HF-X, 1J1P-6PKN-T, 4R7H-7LPC-J, 282N-0WPA-X, 5G35-04K8-E, 366L-2XHV-A, 5L7D-2UGW-T, 4N0L-0FKE-2; Distribution: Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Brazil, Canada, Czech Republic, Dominican Rep, Finland, France, Germany, Greece, Hungary, India, Italy, Japan, Jordan, Netherlands, Norway, Poland, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, United Kingdom, Utd.Arab.Emir. Affected quantity: 146 units (41 US, 105 OUS)
Hazard
A recent software (SW) patch modifies Mobile Event Notification filter settings without providing any indication to the user when upgrading the system.

More from Philips