FDA-Device2016-03-23Class II
Product 68 consists of all product under product code: JDI and same usage: Item no: 787101360 VERSYS CEM/REV/CALCAR 13X 787101361 VERSYS CEM/REV/CALCAR 13X 787101362 VERSYS CEM/REV/CALCAR 13X 787101363 VERSYS CEM/REV/CALCAR 13X 787101560 VERSYS CEM/REV/CALCAR 15X 787101561 VERSYS CEM/REV/CALCAR 15X 787101562 VERSYS CEM/REV/CALCAR 15X 787101563 VERSYS CEM/REV/CALCAR 15X 787101760 VERSYS CEM/REV/CALCAR 17X 787101761 VERSYS CEM/REV/CALCAR 17X 787101761 VERSYS CEM/REV/CALCAR 17X For
Zimmer Manufacturing B.V.
Hazard
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Sold states
US Nationwide Distribution
Affected count
1220
Manufactured in
Turpeaux Industrial Park Rd 1 km 123.4, Mercedita, PR, United States
Products
Product 68 consists of all product under product code: JDI and same usage: Item no: 787101360 VERSYS CEM/REV/CALCAR 13X 787101361 VERSYS CEM/REV/CALCAR 13X 787101362 VERSYS CEM/REV/CALCAR 13X 787101363 VERSYS CEM/REV/CALCAR 13X 787101560 VERSYS CEM/REV/CALCAR 15X 787101561 VERSYS CEM/REV/CALCAR 15X 787101562 VERSYS CEM/REV/CALCAR 15X 787101563 VERSYS CEM/REV/CALCAR 15X 787101760 VERSYS CEM/REV/CALCAR 17X 787101761 VERSYS CEM/REV/CALCAR 17X 787101761 VERSYS CEM/REV/CALCAR 17X For use in total or hemi hip arthroplasty
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1166-2016Don't want to check this manually?