FDA-Device2020-02-19Class II
Ultrasound Transducer (M2736A), that may have been serviced with the following: Pacific Medical Philips Avalon Smart US/TOCO Cable Assembly, NFPHA9020; Pacific Medical Philips M2736A Avalon Smart Ultrasound Bottom Case, NFPHA9515
Pacific Medical Group Inc.
Hazard
Fetal/maternal monitor/monitoring systems serviced or remanufactured using non-OEM equivalent components that have not been appropriately verified or validated. Possible adverse clinical impact: inability to accurately detect and measure fetal heart rate, inability to accurately detect maternal peripheral oxygen saturation, inability to monitor uterine activity to follow contractions, inability to determine temperature, and maternal blood pressure, lack of electrocardiography tracing, patient/clinician burns, and delay in detecting maternal or fetal distress.
Sold states
Worldwide Distribution - US Nationwide in the states of::NH, NY, MN, PA, FL, AZ, LA, GA, IL, HI, CA, OR, SC, OH, NC, IA, OK, AL, MA, MD, WI, MI, AR, WA, SD, NV, KY, TX, MS, DE, NJ, NE, WV, MT, TN, VA, MO, VT, CO, CT, ME, WY, IN, UT, NM, AK, DC, ID, KS, ND, RI, NH. OUS: BEL, ARE, MEX, CAN, CYM, KOR, ZAF, BHR, ESP, ITA, FRA, VUT, AUS, GBR.
Affected count
5897
Manufactured in
212 Avenida Fabricante, San Clemente, CA, United States
Products
Ultrasound Transducer (M2736A), that may have been serviced with the following: Pacific Medical Philips Avalon Smart US/TOCO Cable Assembly, NFPHA9020; Pacific Medical Philips M2736A Avalon Smart Ultrasound Bottom Case, NFPHA9515
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1167-2020Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief