FDA-Device2018-04-04Class II

Sofia 2 Analyzer Catalog Number 20299 Sofia 2 is a bench top analyzer intended to be used with Cassette-based immunofluorescent in vitro diagnostic assays . Sofia 2 is intended for professional and laboratory use.

Quidel Corporation
Hazard

The calibration for the affected batch of analyzers was incorrectly set during the manufacturing process which may lead to a failure when calibration is first performed. There may also be e remote possibility that incorrect qualitative test results for patient samples near the assay cutoff may be reported.

Sold states
US Distribution
Affected count
482 units
Manufactured in
10165 McKellar Ct, San Diego, CA, United States
Products
Sofia 2 Analyzer Catalog Number 20299 Sofia 2 is a bench top analyzer intended to be used with Cassette-based immunofluorescent in vitro diagnostic assays . Sofia 2 is intended for professional and laboratory use.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1183-2018